    题名: 評估體重差異下的假體在PET/CT掃描下的有效劑量
    Evaluating Effective Dose for Different Weight Phantoms undergoing PET/CT Investigations
    作者: 蔡筱涵
    Tsai, Hsiao-Han
    贡献者: 中山醫學大學;醫學科技學院;生物醫學科學學系碩士班;陳健懿
    关键词: 熱發光劑量計;自製研發數學假體;ICRP 60;ICRP 103;AAPM 96;AAPM 204;PET/CT
    Effective doses (E);TLD;self-developed mathematic phantom;PET/CT;ICRP 60;ICRP 103;AAPM 96;AAPM 204
    日期: 2012
    上传时间: 2012-12-21T06:34:57Z (UTC)
    摘要: 正子電腦斷層掃描為準確率極高的檢查項目,廣泛應用於腫瘤病灶的定位及追蹤,並普遍用於健康檢查,其配合螺旋式之電腦斷層,並進行大範圍的全身性掃描,因此劑量及其危險度皆須詳細評估。
    本論文使用熱發光劑量計、侖道假體、以及參考國際輻射單位與度量委員會48號報告和數學亞當假體所設計出的自製研發數學假體,利用林新醫院核子醫學科正子造影中心的西門子正子電腦斷層之biograph 16多切面電腦斷層進行掃描,找出不同體重之假體在進行PET/CT掃描下所得電腦斷層劑量之關係。其暴露條件與實際臨床上應用的條件相同,掃描條件設定為120 kV、Auto care dose 4D和3 mm/slice。
    全身有效劑量計算分為三大部分,第一,由每切片所代表器官之劑量,分別根據國際放射防護委員會60號報告及103號報告進行加權平均,並依其報告制訂之等價劑量公式,計算出全身有效劑量。第二,由工作站的螢幕上所顯示之電腦斷層劑量指標,經由美國醫學物理協會96號報告及204號報告所提供之公式計算出全身有效劑量。以及隨機取26名曾進行PET/CT受檢者之資訊,以AAPM 96及AAPM 204號報告分析。
    本次實驗,10至90公斤假體用60號及103號報告計算結果相差最多至3.5%,體重與劑量關係之方程式分別為mSv=7.14+8.9×10-2×kg及 mSv=6.63+9.9×10-2×kg;AAPM 96及 204報告提出之公式計算之有效劑量因轉換因子及k值的不同而造成其計算結果25%之差異。實際受檢者用AAPM 96及204所計算之有效劑量則分別在6.09至9.49和18.4至23.5 mSv的範圍中。
    經由ICRP 60號報告之危險度評估,經由一次電腦斷層的掃描,10至90公斤假體之全身危險度約為0.037%至0.073%,在安全範圍內。假體體重與劑量相對應關係,所得到之方程式,未來可提供核子醫學科和放射科醫師、放射師臨床診斷之計量的評估參考。
    PET/CT is a kind of high accuracy for functional image modality. It has been widely used in screening cancer for decades. Operation of the PET/CT scanner will be produce huge amount of x-ray scattered from the target, dose caused a significant increase.
    To evaluate effective dose (E) of rando and self-developed mathematic phantom (SDM) derived from ICRU-48 therluminescence dosimeters (TLD-100H) was inserting into these phantoms undergoing PET/CT examinations. All scans, 120kV, auto care dose 4D and 3 mm/slice, the same as patients, were carried out on the 16 slice of BioGraphy PET/CT at Lin Shin Hospital.
    All scans wee performed by senior radiotherapist. E was calculated by (A) in the organs and tissues of interest using TLD and recommended by ICRP 60 and ICRP 103; (B) in DLP method, E was estimated according to CTDIvol that was recorded directly from the console display of PET/CT at the time of the scan and conversion coefficient (k) recommended by AAPM 96 and 204. E was also discussed and compared with 26 volunteers selected in random undergoing PET/CT examinations and those of phantoms.
    Calculated E based on ICRP 60 and 103 was different more than 3.5% among these SDM phantoms herein. The weight and dose relationship of equation are mSv=7.14+8.9×10-2×kg and mSv=6.63+9.9×10-2×kg, respectively. In addition, the conversion factor and k factor created 25% E difference between AAPM 96 and 204 between 10 and 30 kg phantoms. Volunteers’ E were ranging from 6.09 to 9.49 and 18.4 to 23.5 mSv by AAPM 96 and 204, respectively.
    The risk evaluated by ICRP 60 was estimated 0.0037% to 0.073% during one CT scan for these phantoms and found the extra risk was negligible. A simple equation of fit was derived to calculate of E for a patient of any weight that advisory, physician, radiotherapist, for diagnostic reference dose can be assessed.
    URI: https://ir.csmu.edu.tw:8080/ir/handle/310902500/5717